Philips North America Llc — Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Official product recall notices and clearly labeled enforcement reference records.
Reference record — not a live warning · enforcement-reference · official-source
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
- Posted by source
- Source last updated
- Current version retrieved
- Provider
- FDA — device enforcement reference
- Location
- Cambridge, MA, US
- hazard
- The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
- notice
- Periodic enforcement reference; not a real-time safety alert or a reliable recall-lifecycle tracker.
- product
- Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
- distribution
- U.S. and U.S. Territories.
- event number
- 99865
- manufacturer
- Philips North America Llc
- recall number
- Z-3155-2026
- source status
- Open, Classified
- freshness minutes
- 180